DEVELOPMENT AND VALIDATION OF ANLC–MSMSTECHNIQUE FOR THE QUANTITATION OF TEPOTINIBINHUMAN PLASMA
The goal of the study was to develop a sensitive, accurate, and linear LC-MSMS technique for quantifying Tepotinib. Chromatographic separation was attained by a Zorbax ODS/C18 column (50 mm ×4.6mm, 2.1µm) with a solvent system consisting of methyl alcohol, 0.1% HCOOH, and ACN in the proportion of 25:15:60 (v/v/v), deliveredat 0.5mL/min flow. In MRM mode, detection was performed by monitoring ion transitions between m/z 493.17and436.17 for Tepotinib and between m/z 488.2 and 401.1 for the Dasatinib internal standard. The linear equationwasy=0.0026x-0.0078, with an r² values of 0.9994. The %CV findings for matrix effect of LQC and HQClevels were3.63% and 3.91%, correspondingly. Mean recovery percentages for Tepotinib at HQC (525µg/ml), MQC(350µg/ml), and LQC (49 µg/ml) were 102.94%, 96.61%, and 95.27%, correspondingly. The accuracy of theanalytical method, both interday and intraday, was assessed by estimating %CV values. The quality controlsexhibited variability ranging from 2.86% to 4.75% at concentrations of 17.5, 49.0, 350.0, and 525.0 ng/mL. Thedeveloped method was validated in compliance with USFDA guiding principles and demonstrated suitabilityfor determining tepotinib in plasma for kinetics and quality control studies.